60 Multimode Source for 6100 Series Single Quad LC/MS User Guide
3 Reference
Guidelines for multimode source settings in MM-ES+APCI mode
Guidelines for multimode source settings in MM-ES+APCI mode
Drying gas flow Always set to 5 L/min (IR does the drying)
Nebulizer pressure Start at 40 psi. You will use the nebulizer pressure and the
corona current to balance ESI and APCI. A higher nebulizer pressure results
in more ESI. A lower nebulizer pressure results in more APCI.
Drying gas temperature Set to 300°C. Remember you want the vaporizer to
control the drying and to be able to maintain the set point.
Vaporizer Start at 200°C. A higher temperature usually works better for
MM-APCI. MM-ESI is usually unaffected except for sodium adduction. The
vaporizer will compensate for flow rate or solvent composition.
Capillary Voltage Set to 2000 V. A lower voltage works better for fast
switching. You are balancing ESI and APCI response. A setting of 2000 is a
good compromise.
Corona current Set to 1 µA for positive or negative polarity. More corona
current will reduce the ESI signal and increase the APCI signal.
Charging Voltage Set the charging voltage to 2000 V.
NOTE
Click Instrument > MSD Spray Chamber to set the Method Spray Chamber to
MM-ES+APCI before you edit the Set Up MSD Signals dialog box.
NOTE
The above set of parameters is provided as a starting point and should yield good response
for many analytes. To obtain a specific response, i.e., high sensitivity detection of a
compound or class of compounds, or a balanced response to a wide variety of compound
classes, these parameters and the LC mobile phase composition should be examined
further using traditional techniques for optimizing a method.