User guide

40 Multimode Source for 6100 Series Single Quad LC/MS User Guide
3 Reference
Safety
Fluid drained from the multimode chamber is made of solvent and sample
from your analyses. The fluid in the mechanical pump collects traces of
samples and solvents. In addition, unnebulized solvent and sample collects
at the bottom of the spray chamber. Connect the drain on the bottom of the
spray chamber to a closed container.
Handle and dispose of all fluids using precautions appropriate for their
biohazardous and biological content. Comply with local environmental
regulations.
Handle all used pump fluid as hazardous waste. Dispose of used pump fluid
as specified by your local regulations.
Environmental Conditions
This equipment must be installed in an environment of Category II installation
as defined in IEC 664. Check that the supply voltage does not fluctuate more
than +10% or -10% of rated voltage.
Equipment Class Class 1 Laboratory Equipment
Pollution Degree 2
Installation Category II
Environment
Indoor Use
Altitude Not to exceed 2300 m
Electrical supply 100 - 240 V AC, 50/60 Hz, 1.2 A
Mains supply voltage Fluctuations not to exceed 10% of nominal supply
voltage
Operating Temperature 15 to 35°C (59 to 95°F)
Humidity < 90% RH at 35°C
Operational Conditions
If the G1978B is used in a manner not specified by Agilent Technologies, the
protections provided by the G1978B may be impaired.