User guide

Multimode Source for 6100 Series Single Quad LC/MS User Guide 27
Installation 2
Step 2. Install the HV control PCA and cables
Figure 13 Connecting the APCI HV power supply to the valve PCA.
10 Insert one end of the APCI Needle Interlock cable, G1960-60856, through
the slot at the front of the system and then plug it to the APCI HV
connector. Attach the other end to the chassis with the o-ring and the nut.
Figure 14 Connecting the APCI HV to the chassis.
11 Insert the cable, G1960-60858, to the top slot and attach it to the chassis.
Plug the other two ends into the multimode HV PCA.
Figure 15 Connecting the HV PCA to the chassis.
12 Close the AUX Module cover and reconnect all cables.
13 Install the MM source to the system and connect all connectors.