User guide

Multimode Source for 6100 Series Single Quad LC/MS User Guide 25
Installation 2
Step 2. Install the HV control PCA and cables
Step 2. Install the HV control PCA and cables
1 Turn off the system power and remove the system power cord.
The power cord should be kept intact if the vacuum control switch box is
used. The switch box is intended to keep the vacuum on while a CE works
on the electronics. The switch box is for CE use only.
2 Remove the CDS cover, top, side, front, and the Aux Module cover.
3 Disconnect the ribbon cable that connects the valve PCA to the
Vcap/Vchamber power supply. Then disconnect the Vcap and Vchamber
cable from the power supply.
Figure 9 Disconnecting the Vcap/Vchamber power supply from the valve PCA (left) and the Vcap/Vchamber.
4 Place the MM HV power supply PCA in the slot between the valve PCA and
the Vcap/Vchamber. Secure the board by pressing it down into its slot and
then attach it with two screws.
5 Connect the short gray cable from valve PCA to the multimode HV power
supply.
Figure 10 Connecting the valve PCA to the multimode HV power supply.