User guide
20 Multimode Source for 6100 Series Single Quad LC/MS User Guide
1 Basic Operation and Maintenance
To clean the multimode source weekly
6 Remove the spray shield. Use abrasive paper to gently clean the end of the
capillary cap.
7 Dampen a clean cloth and wipe the end of the capillary cap.
8 Reinstall the spray shield.
9 Use abrasive paper to gently clean the spray shield. Dampen a clean cloth
and wipe the spray shield.
10 Rinse the area around the spray shield. Then wipe the area around the
spray shield.
11 Reinstall the spray chamber on the instrument.
12 Replace the nebulizer and APCI corona needle.
13 Install the thermocouple probe and adjust it so that it protrudes 15 mm
from the inner spray chamber wall. See Figure 5 on page 17.
14 Replace the cosmetic cover.
15 Close the spray chamber.