User guide

Multimode Source for 6100 Series Single Quad LC/MS User Guide 19
Basic Operation and Maintenance 1
To clean the multimode source weekly
To clean the multimode source weekly
The cleaning procedure for cleaning the multimode source weekly is similar to
the procedure for cleaning the source daily. The main difference is that the
multimode source is removed from the instrument.
Before you begin, check that you have:
Abrasive paper, 8000 grit (p/n 8660-0852)
Cloths, clean, lint-free (p/n 05980-60051)
Cotton swabs (p/n 5080-5400)
Gloves
Mobile phase from the current method, or clean isopropanol, reagent
grade or better
Wash bottle
Water, reagent-grade or better
1 Do the steps in “To remove the multimode source” on page 29.
2 Fill the spray chamber with clean mobile phase, or with a mixture of
isopropanol and water.
3 Scrub the corona insulator and the interior of the spray chamber with a
clean cotton swab.
4 Empty the spray chamber.
5 Wipe the interior of the spray chamber with a clean, lint-free cloth.
NOTE
Recent residue should be soluble in the mobile phase. If you are not sure what mobile
phase was used recently, a mixture of 50% isopropanol and 50% water works well as a
general cleaning solution.
WARNING
Some mobile phases are hazardous chemicals. Use caution that is appropriate for
the current mobile phase.
WARNING
The inside of the spray chamber has sharp edges, such as on the separator. Pay
close attention when you wipe the interior of the spray chamber.