To clean the multimode source daily
To clean the multimode source daily
You should clean the multimode source daily or any time you suspect
carryover (contamination) from one sample or analysis to another. The
multimode source spray chamber is made of 316 stainless steal which is the
same material the spray shield is made of.
Before you begin, check that you have:
• Abrasive paper, 8000 grit (p/n 8660-0852)
• Cloths, clean, lint-free (p/n 05980-60051)
• Cotton swabs (p/n 5080-5400)
• Gloves
• Mobile phase from the current method or clean isopropanol, reagent
grade or better
• Wash bottle
• Water, reagent-grade or better
1 Turn off the spray chamber. Use the Spray Chamber menu item in Tune
and set all gas flows and temperatures to minimum values of zero.
2 Remover the nebulizer and APCI corona needle.
3 Remove the cosmetic cover. The thermocouple probe will need to be
removed if any wiping of the spray chamber will be done. Then, open the
spray chamber.
4 Rinse the interior of the spray chamber using the wash bottle filled with the
current mobile phase or with a mixture of isopropanol and water.
.
WARNING
Do not touch the multimode source or the capillary cap. They may be very hot. Let the
parts cool before you handle them.
NOTE
Recent residue should be soluble in the mobile phase. If you are not sure what mobile
phase was used recently, a mixture of 50% isopropanol and 50% water works well as a
general cleaning solution.