Technical data
28 Multimode Source for 6400 Series Triple Quad LC/MS Set-Up Guide
2 Performance Verification
To set up a method to use the multimode source
To set up a method to use the multimode source
1 In the MassHunter software, change the Context to Acquisition.
2 In the MS QQQ tab, set Ion source to MMI (see Figure 22 on page 29).
3 In the Time Segments table, chose an ionization mode from the Ion Mode
list. You may set the ionization mode to one of the following:
• ESI
• APCI
• ESI+APCI
The Ion Mode selection ESI + APCI will specify a method for Simultaneous
ESI and APCI operation.
Note that the Ion Mode selection is only visible if Ion source is set to MMI.
4 In the Source tab, set the desired source conditions. See “Guidelines” in the
Agilent G1978A/B Multimode Source Maintenance Guide for suggested
source conditions for the multimode source for the different ionization
modes.
5 Make any other changes that are necessary for your method.
6 Save the method.