Technical data

Multimode Source for 6400 Series Triple Quad LC/MS Set-Up Guide 27
Performance Verification 2
To autotune with the multimode source
To autotune with the multimode source
1 Remove the G1948B Electrospray source and install the G1978B multimode
source.
2 Pour the Electrospray or APCI calibrant back into its original bottle or
another suitable container, rinse the calibrant bottle with acetonitrile, pour
the MMI-L Low Concentration Tuning Mix (G1969-85020) calibrant into the
calibrant bottle, and attach the calibrant bottle back onto the CDS.
3 Set the Context view to Tune in the MassHunter Workstation program.
4 Load an autotune file that was generated with the G1948B Electrospray
source.
5 Run an Autotune using the G1978B multimode source.