Technical data

26 Multimode Source for 6400 Series Triple Quad LC/MS Set-Up Guide
2 Performance Verification
To do sensitivity verification
4 Print each signal-to-noise calculation report for each solvent blank and
reserpine injection. Include the chromatogram in the printout.
5 Open the Excel spreadsheet D:\MassHunter\Support\Checkout\Sensitivity
Checkout Report.xls. Fill in the values to calculate the average signal to noise
and save the spreadsheet.
The average signal-to-noise must be greater than 50:1.
6 Print the Excel signal-to-noise report.