Technical data

24 Multimode Source for 6400 Series Triple Quad LC/MS Set-Up Guide
2 Performance Verification
To do sensitivity verification
To do sensitivity verification
Figure 19 Reserpine checkout worklist
1 Review the worklist to be sure that the method and data paths are correct,
and that the data file names given in the worklist are unique and have not
already been acquired.
2 Run the worklist.
3 When the worklist is finished, calculate signal-to-noise for each injection:
a Load each solvent blank and reserpine sample data file into the
Qualitative Analysis program.
b Generate Extracted Ion Chromatograms of the 195.1 ion.