Technical data
12 Multimode Source for 6400 Series Triple Quad LC/MS Set-Up Guide
1 Installation
Step 2. Install the HV control PCA and cables
11 Insert the cable, G1960-60858, to the top slot and attach it to the chassis.
Plug the other two ends into the multimode HV PCA.
Figure 9 Connecting the HV PCA to the chassis.
12 Close the AUX Module cover and reconnect all cables.
13 Install the multimode source onto the system and connect all connectors.
Figure 10 Installing the multimode source (left) and connecting all connectors.
14 Put back the side, top, front and CDS cover.
15 Plug the system power cord back on and turn the front switch on.
The pump down process will start.
16 Start the MassHunter Workstation program and verify that the software
recognizes the source.
17 Set the Context view to Tune, and in Manual Tune, verify that the system
can generate the proper tune peaks.