3 Installation Verification
Step 9. Run each of the methods created
42 Multimode Source for 6510 Q-TOF LC/MS Set-Up Guide
Step 9. Run each of the methods created
1 Run each of the methods that you just created.
The real time plot below shows the six runs.
2 View the data from Analyst for MM_ESI_pos.wif. Exstract Ion 372- 372.4.
Record peak height Example: 91,000.
ESI +APCI NEG
1-Hexanesulfonic
acid &
9-Phenanthrol
ESI POS
Crystal Violet
ESI NEG
1-Hexanesulfonic
acid
APCI POS
Carbazole
APCI NEG
9-Phenanthrol
ESI +APCI POS Crystal
Violet & Carbazole