Installation Verification 3
Step 6. Create MMCHECKTOF_CI_NEG.m
Multimode Source for 6510 Q-TOF LC/MS Set-Up Guide 37
Ionization Mode MM-APCI
Polarity Negative
1100 Binary Pump 1
Control
Column flow 0.400 mL/min
Stop Time No Limit
Post Time Off
Solvents
Solvent A 100.0 % (65%MeOH:35%H
2
O + 0.2%acetic acid))
Solvent B 0.0 %
Pressure Limits
Minimum Pressure 0 bar
Maximum Pressure 400 bar
Spray Chamber
[MSZones]
Gas Temp 350 °C Maximum 350 °C
Vaporizer 200 °C Maximum 250 °C
Drying Gas 5.0 L/min Maximum 13.0 L/min
Neb Pres 20 psig Maximum 60 psig
VCap (Positive) 1000 V
VCap (Negative) 1000 V
VCharge (Positive) 2000 V
VCharge (Negative) 2000 V
Corona (Positive) 6.0 μA
Corona (Negative) 6.0 μA