Installation Verification 3
Step 2. Set up method names and parameters
Multimode Source for 6510 Q-TOF LC/MS Set-Up Guide 29
Step 2. Set up method names and parameters
1 Create six methods from Default.m for the multimode ESI + APCI LC Demo
Sample (p/n G1978-85000), using these method names:
• MMCHECKTOF_EI_POS.m
• MMCHECKTOF_EI_NEG.m
• MMCHECKTOF_CI_POS.m
• MMCHECKTOF_CI_NEG.m
• MMCHECKTOF_MX_EI POS_CI POS.m
• MMCHECKTOF_MX_EI NEG_CI NEG.m
2 Use these parameters for each method:
Table 1
Parameter/Tab Value
Column Cartridge Hardware, Rapid Resolution, (p/n 820555-901)
SB-C18 Rapid Res 3.5um,2.1x30mm, (p/n 873700-902)
Sample Tab Name: MM Demo Sample
Position 1
Run Type: Standard
Acquisition only
Path D:\PE Sciex Data\Projects\Data
ALS Tab Standard Injection 1µL
Bin Pump Tab:
Flow .4mL/min
Stop time: 3 min
Solvent A 100.0 % (65%MeOH:35%H
2
O + 0.2%acetic acid)
Run time same as pump
Data files (data files
for B.01.03 or greater
use the suffix .d)
Data File: MM_ESI_POS.wiff
Data File: MM_ESI_NEG.wiff
Data File: MM_APCI_POS.wiff
Data File: MM_APCI_NEG.wiff
Data File: MM_ESI_APCI_POS.wiff
Data File: MM_ESI_APCI_NEG.wiff