Technical data

3 Installation Verification
Step 1. Auto tune
28 Multimode Source for 6510 Q-TOF LC/MS Set-Up Guide
Step 1. Auto tune
This step applies to MassHunter Workstation Software - Acquisition for
TOF/Q-TOF revision B.01.03 or higher.
Run autotune with the G1969-85000 ESI-L Low Concentration Tuning Mix.
There are no tune specific methods.
Tune the 6220 in 2GHz extended dynamic range for both positive and
negative.
Tune the 6210 in Standard (3200 m/z) mode 1GHz.