Technical data
Table Of Contents
- Content
- Installation
- Set-Up
- Installation Verification
- Step 1. Auto tune
- Step 2. Set up method names and parameters
- Step 3. Create MMCHECKTOF_EI_POS.m
- Step 4. Create MMCHECKTOF_EI_NEG.m
- Step 5. Create MMCHECKTOF_CI_POS.m
- Step 6. Create MMCHECKTOF_CI_NEG.m
- Step 7. Create MMCHECKTOF_MX_EI POS_CI POS.m
- Step 8. Create MMCHECKTOF_MX_EI NEG_CI NEG.m
- Step 9. Run each of the methods created
- Step 10. Calculate the response of Multimode Demo
- Step 11. Fill out Multimode Report for calculation of peak heights
- Index
3 Installation Verification
Step 1. Auto tune
28 Multimode Source for 6510 Q-TOF LC/MS Set-Up Guide
Step 1. Auto tune
This step applies to MassHunter Workstation Software - Acquisition for
TOF/Q-TOF revision B.01.03 or higher.
• Run autotune with the G1969-85000 ESI-L Low Concentration Tuning Mix.
There are no tune specific methods.
• Tune the 6220 in 2GHz extended dynamic range for both positive and
negative.
• Tune the 6210 in Standard (3200 m/z) mode 1GHz.