Technical data

Multimode Source for 6510 Q-TOF LC/MS Set-Up Guide 17
Installation 1
To convert from ESI or APCI to the multimode source
Figure 13 Multimode spray shield
11 Screw the multimode spray shield into the holder for the spray shield. See
Figure 14.
Figure 14 Multimode spray shield installed
12 Remove the shipping cover from the multimode source spray chamber.
Field shaping electrode
9 o'clock position
Field shaping electrode
6 o'clock position
NOTE
The field shaping electrodes should be in the nine o’clock and the six o’clock position.
Loosen the end plate screws on each side to adjust the field shaping electrodes position.