Technical data

Agilent Signal Generators Programming Guide 45
Using IO Interfaces
Using LAN
2. At the command prompt, type in telnet.
3. Press the Enter key. The Telnet display screen will be displayed.
4. Click on the Connect menu then select Remote System. A connection form (Figure 2- 5) is displayed.
Figure 2-5 Connect Form (Agilent IO Library version J.01.0100)
5. Enter the hostname, port number, and Ter mTy pe then click Connect.
•Host NameIP address or hostname
•Port5023
•Term Typevt100
6. At the SCPI> prompt, enter SCPI commands. Refer to Figure 2- 6 on page 46.
7. To signal device clear, press Ctrl–C on your keyboard.
8. Select Exit from the Connect menu and type exit at the command prompt to end the Telnet
session.
Using Telnet On a PC With a Host/Port Setting Menu GUI
1. On your PC, click Start > Run.
2. Type telnet then click the OK button. The Telnet connection screen will be displayed.
3. Click on the Connect menu then select Remote System. A connection form is displayed. See Figure
2- 5.
4. Enter the hostname, port number, and Ter mTy pe then click Connect.
•Host Namesignal generator’s IP address or hostname
•Port5023
•Term Typevt100
5. At the SCPI> prompt, enter SCPI commands. Refer to Figure 2- 6 on page 46.
6. To signal device clear, press Ctrl–C.
7. Select Exit from the Connect menu to end the Telnet session.