Technical data
Agilent Signal Generators Programming Guide 201
Programming the Status Register System
Status Groups
Data Questionable BERT Status Group
NOTE This status group applies only to the E4438C with Option UN7. Refer to Table 4- 13 on
page 202 for more information.
The Data Questionable BERT Status Group is used to determine the specific event that set bit 12 in
the Data Questionable Condition Register. The Data Questionable Status group consists of the Data
Questionable BERT Condition Register, the Data Questionable BERT Transition Filters (negative and
positive), the Data Questionable BERT Event Register, and the Data Questionable BERT Event Enable
Register.