Understanding the Extended Measurements Report
3-20 Understanding ECG Analysis and the Predictive Instruments Applications
Global Rhythm Parameters
The following parameters provide global information for beats in the ECG.
Table 3-6: Global Rhythm Parameters
Parameter Units or Value Description
Atrial Rate Beats per minute The representative atrial rate for the analysis
interval. This is not a simple arithmetic average.
Low Ventr Rate Beats per minute The lowest ventricular rate during the analysis
interval.
Mean Ventr Rate Beats per minute The average ventricular rate during the analysis
interval.
High Ventr Rate Beats per minute The highest ventricular rate during the analysis
interval.
Flut-Fib Indicator N/A Indicates approximate number of flutter-like or
coarse fibrillatory waves per lead.
Fixed Mult P Morph Yes or No Indicates that all P waves are of consistent mor-
phology.
Mult P Test Valid Yes or No Indicates that the tests performed to detect mul-
tiple P waves produced consistent results.
Delta Wave Count N/A Number of QRS complexes with pronounced
delta waves.
Delta Wave% Percentage Percent of total beats with pronounced delta
waves.
Bigeminy Count N/A Total number of beats in a bigeminy pattern,
whether or not they are contiguous.
Bigeminy String N/A Total number of beats in the longest continuous
bigeminy pattern.
Trigeminy Count N/A Total number of beats in a trigeminy pattern,
whether or not they are contiguous.