User`s guide

Understanding the Printed Report
PageWriter 200/300pi M1771A/1770A Cardiograph 2-15
2
Understanding the Printed Report
The PageWriter 200/300pi provides Auto, Manual, and Extended
Measurements reports. This section describes the Manual and Auto reports.
The Manual report features closely resemble those of the Auto report. The
Extended Measurements report is described in “Understanding ECG Analysis
and the Predictive Instruments Applications”.
Choosing a Report Format
An Auto report prints a one- or two-page summary of all 12 cardiograph
leads. (More pages may be needed to print additional interpretation
statements. See Table 2-4). A Manual report prints continuously until the
key is pressed. The Auto and Manual report formats are available for
both standard and Cabrera leads. The Cabrera lead order is an alternative limb
lead order in which aVR is inverted and shown as -aVR. Lead order is aVL, I,
-aVR, II, aVF, III, V1 through V6. The Cabrera lead order makes it easier to
visualize waveform progression in the frontal plane.
Using the softkeys below the cardiograph’s display, the user can select the
desired report format and lead configuration. To change to a different report
format, see “Changing the Report Format” later in this section.
Table 2-4: Auto Report Length Configurations
Number of pages in Auto report
Model At 25 mm/sec At 50 mm/sec
300pi with
Interpretation off
12
300pi with
interpretation on
1 or more 2 or more
200 1 2