User`s guide
Recording a Manual ECG
PageWriter 200/300pi M1771A/1770A Cardiograph 2-11
2
ID, Continue? is displayed. Press to record the ECG
without a patient ID. Press to enter patient ID information.
l If you did not enter a patient ID in the above step but have entered
patient ID information since the cardiograph was turned on, the
question
Continue using Old ID? is displayed.
If you have entered a patient identification number and want to take
more ECGs from the same patient, press .
If you are recording an ECG from another patient, press
or .
NOTE Your cardiograph automatically checks for a patient ID each time you start an
ECG, unless it is configured for no patient ID information. To bypass this
check and use the last entered patient ID, press a second time.
6. The cardiograph prints the ECG continuously until you press the
key.
NOTE If accurate ECG ST contours are required for ECGs recorded in Manual
mode, do not use the 0.5 Hz baseline wander filter. This filter suppresses
baseline wander to the extent that it may alter the ST segment. Instead,
configure your cardiograph to use the 0.15 Hz or 0.05 Hz baseline wander
filter. Regardless of the filter used, the rhythm characteristics of the ECG are
accurately recorded. Refer to Configuring Your Cardiograph for details on
configuring filters.
The leads recorded by the manual ECG, the chart speed, sensitivity, and filter
status can be changed at any time, even while a manual ECG is running. To
change these settings simply select , , , , or
keys, as appropriate. Note, however, that the ECG printer will pause as the
cardiograph changes its recording status.
l The key sequence is: 3 ⇒ 6 ⇒ 12 ⇒ 3 ...
Ye s
No, Enter ID
Ye s
No, Edit Old ID
No, Enter New ID
0DQXDO
6WRS
Format Leads Speed Size
)LOWHU
Format