NOTE The equipotential connector is used when the cardiograph must be plugged
into an ungrounded outlet. See Patient and Operational Safety Notes in Get-
ting Acquainted for more information about using the equipotential connector.
The Battery
Your cardiograph requires the battery to be installed for proper operation
even if the cardiograph is plugged into AC power. The cardiograph cannot
operate without the battery.
Use only part number M2460A batteries in the cardiograph.
Installing the Battery
To Install the Battery:
NOTE. Do not remove the shrinkwrap surrounding the battery.
1. Make sure the cardiograph is unplugged from AC power.
2. Turn the cardiograph bottom-side up.
3. Slide the battery door in the direction of the arrow shown in Figure A-
until it unlatches (approximately 1/2 inch). Lift off the door.
4. Install the new battery in the battery compartment as shown in Figure A-
and plug the battery connector into the cardiograph.
5. Place the battery door into its slots and slide the door in the opposite
direction of the arrow shown in Figure A-2 until it latches.
6. Turn the cardiograph top-side up.
7. Plug the cardiograph into AC power.
8. Check that the AC indicator light is on. The unit is now in Standby (off)
mode with the battery charging.
After you finish setting up the cardiograph, it may be used on a limited basis
until the battery is fully charged. Charging the battery as soon as possible for