User`s guide

Error Messages
PageWriter 200/300pi M1771A/1770A Cardiograph 6-5
6
3. Press or . If the cardiograph turns itself to Standby (off), the
battery is not operating properly.
If the cardiograph still won’t operate, call your local Agilent Technolgies ser-
vice representative.
If the Cardiograph Won’t Print a Manual Report
l Is the paper sensor lens dirty or obstructed?
Clean the paper sensor lens. See “Maintaining the Cardiograph” for
more information.
l Is the cardiograph out of paper?
Load paper. See “Loading the Paper” in “Maintaining the Cardio-
graph” .
Error Messages
The error messages that display on the screen will instruct you as to what
action to take. If it is something that you can correct, the message will instruct
you what to do. If an error number displays, perform the following steps:
1. Turn the cardiograph to Standby (off) from the front panel.
2. Wait 20 seconds or more and then turn the cardiograph on again.
$XWR