Connecting the Printer
A-6
4. Make sure that the infrared adapter/interface is positioned within the 30°-wide
range of the cardiograph’s infrared port.
5. Turn on the printer and the infrared adapter.
6. Turn on the cardiograph and start recording an ECG. Allow 1.5 minutes for the
printer to print out an ECG report. The indicator on the infrared adapter/inter-
face lights up when the printer receives data from the cardiograph.
Please refer to the relevant printer manuals for detailed instructions about how to
operate the printer.
Use only black and white cartridges. The color cartridge does not work with the
M2662A cardiographs.
The printer and its accessories (e.g. the infrared to parallel converter) may
not meet the leakage current requirements specified in IEC 60601-1 safety
standards. According to IEC 60601-1-1 safety standard, do not place the
printer and its accessories within 1.5 m (4.5 ft) of the patient.
NOTE
NOTE
WARNING