8Removing the Battery
To remove the battery:
1. Make sure the cardiograph is turned off and disconnected from AC power.
2. Turn the cardiograph bottom-side up.
3. Loosen the screws on the battery door. Lift off the door, as shown in the
following figure.
Figure A-2 Removing the Battery Door
4. Unplug the battery connector from the cardiograph by squeezing the edges of
the connector and pulling it straight out.
5. Remove the battery and cable.
6. If the battery has been removed for storage, place the battery cover into its slots
and tighten the screw.
After replacing a battery, you should check the date and time configuration of your
cardiograph. When the cardiograph is without power for about 20 seconds, the
cardiograph may lose the date and time information and you have to re-configure it.
Properly dispose of or recycle depleted batteries according to local
regulations. Do not disassemble, puncture or incinerate the depleted
batteries.