User manual

Notice
viii Revision 0.1
The printer and its accessories (e.g. the infrared to parallel converter) may
not meet the leakage current requirements specified in IEC 60601-1 safety
standards. According to IEC 60601-1-1 safety standard, do not place the
printer and its accessories within 1.5 m (4.5 ft.) of the patient.
If you use the cardiograph during defibrillation, please check the electrode
instructions first. The cardiograph recovery time after defibrillation
specified in IEC 60601-2-25 is affected by the type of electrodes used. Use
of conductive gels with the electrodes improves ECG recovery times after
defibrillation.
The ECG recovery time for defibrillation does not meet the requirements of
IEC 60601-2-25 when using reusable electrodes. You can achieve the ECG
recovery time required by IEC 60601-2-25 with some disposable electrodes
by setting the cardiograph to 0.5Hz Baseline Wander filter setting.
The AC/DC adapter supplied with this cardiograph is an integral part of the
cardiograph’s safety features. Using any other AC/DC adapter may
compromise the cardiograph’s safety as well as performance.
[US] Federal law restricts this device to sale by or on the order of a physician.
Use of accessories other than those recommended by Agilent Technologies
may compromise product performance.
If non-Agilent recommended data transmission cable is used, the cardio-
graph may not meet the Radiated Emission Standard found in CISPR 11.
WARNING
WARNING
WARNING
CAUTION
WARNING
WARNING