User manual

Error Messages
7-4
7Error Messages
The error messages that display on the screen will instruct you as to what action to
take. If it is something that you can correct, the message will instruct you what to
do. If an error number displays, perform the following steps:
1. Turn the cardiograph off from the front panel.
2. Wait 20 seconds or more and then turn the cardiograph on again.
7Identifying Printing Problems
The following table shows symptoms and solutions to problems that can occur
when printing an ECG. Refer to the printer documentation for troubleshooting of
other printing problems.
"IrDA Link Disrupted" messages may occur with power line transients as low as
200 V.
Table 7-3 Printing Problems and Solutions
Message Likely Cause Possible Solutions
Incompatible printer Incompatible printer used. Use the recommended printer.
Printer Not Detected Printer or infrared adapter not
on.
Printer not connected properly.
Infrared interface not ready.
Turn on the printer and infrared
adapter.
Connect the printer properly.
Turn off the cardiograph, wait
10 seconds and turn it on again.
IrDA Link Disrupted Infrared communications
stopped.
Remove any partition between
the printer and the cardiograph
and print again.
NOTE