User manual
6-1
Chapter 6 Transmitting
and Faxing Auto ECGs
6 Transmitting and Faxing Auto ECGs
5Transmitting ECGs
The cardiograph can transmit Auto ECGs to the Agilent M1765A ECG Manager
and to a Group III facsimile machine.
You cannot transmit Rhythm ECGs. You can only transmit Auto ECGs that have
been stored.
Three typical ECG transmission situations are:
• An ECG transmitted from the cardiograph at the bedside to the Agilent M1765A
ECG Manager for printing, overreading and storing.
• ECGs recorded on rounds and then transmitted to the Agilent M1765A ECG Man-
ager in another area of the institution.
• An ECG sent to another institution for overreading or further analysis.
Equipment connected to the cardiograph’s RS-232 connector can cause
ground leakage current exceeding the maximum specified in IEC60601-1
safety standards. Do not connect any equipment to the RS-232 connector
during cardiograph operation when the patient cable is connected to a
patient.
If non-Agilent recommended data transmission cable is used, the cardio-
graph may not meet the Radiated Emission Standard found in CISPR 11.
NOTE
WARNING