Specifications

$JLOHQW0$996HULHV6HUYLFH*XLGH
7HVWLQJ0DLQWDLQLQJWKH$JLOHQW996HULHV
Testing & Maintaining
the Agilent V24/V26
Series
3HUIRUPLQJDQ((35208SJUDGH
You may use this function when repairing a failed EEPROM, replacing a board, or for future
upgrades.
6WHS From the Instrument Configuration screen, press the softkey
to display the EEPROM Upgrade main screen.
6WHS Press the softkey to display the Select EEPROM screen
shown in Figure 2-7.
6WHS Press again to select the EEPROM upgrade (indicated on the
screen after the message Upgrade will be done to:)
6WHS Complete one of the following steps:
a. To perform an EEPROM upgrade ZLWKRXWVDYLQJFXVWRPHUFRQILJXUDWLRQVHWWLQJV,
press .
b. To perform an EEPROM upgrade DQGVDYHFXVWRPHUFRQILJXUDWLRQVHWWLQJV, press
.
6WHS Press to begin the programming process. The system displays the mes-
sage EEPROM programming in process. When the process is complete
(approximately one minute) the system displays the message... Programming
done.
6WHS Press to return to the Instrument Configuration screen.
6WHS Power down the monitor, disconnect the Flash Programming Tool and reinstall the
service port connector cover.
6WHS Power up the monitor and erase the Status Log. (Press ,
, , and .)
6WHS Perform the Performance Assurance test indicated in this chapter.
1RWH7KHIROORZLQJ((35208SJUDGHRU&ORQLQJ6FUHHQILJXUHVDUHIRULOOXVWUDWLRQ
SXUSRVHVRQO\DQGGRQRWVKRZXSJUDGHVWRWKHODWHVWUHOHDVH$
((35208SJUDGH
6HOHFW((3520
6HOHFW((3520
3UJ)DFW'HIDXOWV
3UJ8VHU6HWWLQJV
Confirm
Monitor Setup
Monitor Setup
0RUH&KRLFHV 6WDWXV/RJ (UDVH6WDW/RJ