Test and Inspection Requirements
This section is intended primarily for Agilent customer engineers and lists the test
requirements for servicing, installation and preventive maintenance. For some tests it
is required that the Agilent engineer documents the results in a service record.
Table 1-1 “When to Perform Test Blocks” explains when to perform the procedures.
Table 1-2 “Test and Inspection Requirements” explains the procedures to be
performed. Table 1-3 “Test and Inspection Requirements – Safety” explains the
procedures to be performed and gives full details of the Safety Tests.
Note: Only Agilent service engineers are required to report the results of the
following safety and performance tests back to the factory. Of the performance tests,
only those for NBP and Sidestream CO
2
need to be reported.
7DEOH:KHQWR3HUIRUP7HVW%ORFNV
6HUYLFH(YHQW
±:KHQWR3HUIRUP
7HVW%ORFNV5HTXLUHG
&RPSOHWHWKHVHWHVWV
Installation of M1205A with NO
System (SDN) or with Medical
equipment connections (ex. AGM or
Vuelink)
Perform Visual and Power On test blocks (see page 2-5).
Installation of M1205A that is
connected to a System (SDN)
Perform Visual and Power On test blocks. If the 78581A safety
test (see the 78581A service manual) has passed previously no
further safety tests are required. If the 78581A safety test has
not been performed, it must be completed.
Installation of monitor that has RS232
connections to non-medical devices
(for example, Laser Printer or PC)
Perform Visual and Power On test blocks (see page 2-5).
Perform Safety (1) test block (see page 2-11).
Repairs where the power supply or
rear housing assembly is replaced
Perform Power On and Safety (2) test blocks (see page 2-11).
Repairs where the monitor has been
dropped
Perform Power On and Safety (2) and (3) test blocks (see page
2-12).
All other M1205A repairs, Hardware
or Software Upgrades
Perform Power On test block.
Preventive Maintenance
1
1.
For Germany only: For modules that do not have the CE mark, contact the Customer Response Center in
Ratingen for information.
Perform the NBP and Sidestream CO
2
Module Performance
test blocks every year (see page 2-6).
For the BIS module no periodic preventive maintenance is
required.
Performance Assurance and Safety
Tests
2
2.
For theBIS module performance and safety testing is optional (Agilent recommends yearly testing).
However, safety testing (Safety test block (3)) is required if there is suspected liquid ingress or after
everytime the case of any of the BIS components has been opened.
Perform all other Performance and Safety test blocks (1), (2)
and (3) once every two years.