Specifications
$JLOHQW0$996HULHV6HUYLFH*XLGH
5HSODFHPHQW3DUWV
Replacement Parts
1
Front Housing Kit does not include Flat Panel Display Assembly.
2
Insulator attaches to Board Shield to insulate Flat Panel Display Cable (old insulator not reusable).
3
RFI shielding for external defib sync cable.
4
Color Flat Panel bracket includes RFI springs.
5
When ordering the color display controller (DSPC Bd.), bracket with RFI springs used for CMS must be
removed (2 screws) prior to installation into Model V24CT/V26CT. Bracket must be installed onto defective
DSPC Bd. before returning to Agilent.
6
If you are replacing the flatscreen controller card with Asian/Eastern Europe language option fitted, you need to
set the DIP switches on the function card to select the correct language:
/DQJXDJH ',36ZLWFK6HWWLQJV
Factory Default
a
a. supports all languages except Japanese, Chinese (PRC), Taiwanese (ROC), Greek, Polish, Russian and Czech
closed closed closed closed
Japanese open open open open
Chinese (PRC) closed open open open
Taiwanese
(ROC)
open closed open open
Greek, Russian,
Polish, Czech
open open closed open
Factory Default
Japanese
Chinese (PRC)
Taiwanese (ROC)
Greek, Russian,
Polish, Czech