The following table depicts the slight Power Supply output voltage differences between the
monochrome Model V24, color Model V24C/V26C, and color battery Model V24/V26.
3RZHU6XSSO\,QSXW9ROWDJH
The power supply uses the following input voltages:
AC Input: 90 to 250 VAC continuous, 48 to 66 Hz
Maximum Power Consumption: 150 Volt-Amperes
%DWWHU\3RZHU6XSSO\
Capacity: At least 1 hour on full charge at 25°C (77°F) with two batteries. For example, the
discharge time for the module configuration option #A24 (with ECG/Resp, NBP, SpO
2
,
Pressure, Recorder) is 1 hour. The discharge time for the module configuration option #A22
(with ECG/Resp, NBP, SpO
2
, Pressure) is 1 hour 15 minutes.
Charging time: 4 hours to 90% of full capacity if monitor is off. 16 hours to 90% of full
capacity if monitor is on.
If the Agilent Battery Charger (M1278A) is used, the charging time is 2.5 hours to 90%
capacity.
Monochrome Model V24 Color Model V24C/V26C Color Model V24CT/V26CT
+5 Volts @ 2.0 Amps, max +5 Volts @ 2.5 Amps, max +5 Volts @ 2.8 Amps, max
+12 Volts @ 2.0 Amps, max +12 Volts @ 0.75 Amps, max +12 Volts @ 0.65 Amps, max
+60 Volts @ 0.35 Amps, max +60 Volts @ 0.35 Amps, max +60 Volts @ 0.3 Amps, max