Repairing the Agilent
Model V24CT/V26CT
5HPRYLQJWKH3DUDPHWHU0RGXOH5DFN
Press the power switch to OFF and disconnect the AC power cord.
Remove Lead Acid batteries.
Remove parameter modules from parameter module rack by pressing up on retaining tab
and sliding parameter modules from rack.
If you have a 6-slot rack docked to the V24CT/V26CT, press and hold retaining clip for-
ward. Use other hand to hold monitor steady. Pull the parameter module rack from the
monitor.
Front Housing Removal Procedures
For the rest of the front housing removal procedures see Chapter 9.