Repairing the Agilent
Model V24
Removal Procedures
This section presents procedures to disassemble the instrument to the extent required to
remove and replace faulty assemblies. Do not further disassemble the product past the point
described in these procedures.
High Voltage - Voltages dangerous to life are present in the instrument. Do not perform any
disassembly procedures (other than parameter module removal) with power applied to the
instrument. Failure to adhere to this warning could cause serious injury or death.
Before doing any disassembly, turn power off and disconnect the Local Distribution Cable,
AC power cable, Plug-in Satellite Rack cable, Defib sync cable, and RS232 Cable (where
appropriate).
7RROV5HTXLUHG
• small flat-bladed screwdriver
• small jeweler-sized Phillips screwdriver
• Pozidrive screwdriver (medium)
:$51,1*
TO MAIN
TO SCC
DARK COLOR
LIGHT COLOR
OPEN
CONNECTOR DOOR
LOCAL
DISTRIBUTION
CABLE
PLUG-IN
SATELLITE
RACK
CABLE
POWER
PARAMETER MODULES
9 Pin
25 Pin
RS232 Cable
To local printer, AGM or computer
Note: Cable adapters may be
required depending on device
being connected.
(optional)