Specifications

Troubleshooting Plug-in Modules 5-27
$JLOHQW0$996HULHV6HUYLFH*XLGH
Troubleshooting Plug-
in Modules
Incompatible
Language
BLOOD-
ANL CON-
FIG
297 V24/V26 language and
module language are
not compatible.
Call your service repre-
sentative.
Module Error,
Use Electronic
Simulator
Blood Analy-
sis failed
48, 50-52,
58-62, 70
A cartridge or elec-
tronic simulator was
inserted at an angle or
the simulator is mal-
functioning.
If the error occurs when
using a cartridge, run the
electronic simulator.
If the error occurs when
running the simulator, use
another simulator.
If the error recurs, the
module or simulator may
need repair - call your
service representative.
Module Error,
See Manual
BLOOD-
ANL EQU.
MALF
1, 23, 63,
65-68,
72-74,
76-78, 82,
85, 86,
89, 90,
91-94,
256-296,
298-299
May have various mal-
function causes located
in the module.
Use an electronic simula-
tor twice then use a car-
tridge with sample or
control solution.
If an error condition
occurs, call your service
representative.
Cartridge Not
Inserted Properly
Blood Analy-
sis failed
47 Cartridge or simulator
is probably not pushed
in completely.
Cartridge was removed
after starting a mea-
surement.
Reinsert the cartridge or
electronic simulator.
Start measurement with a
new cartridge.
If error recurs, call your
service representative.
Cartridge Preb-
urst, Use Other
Cartridge
Blood Analy-
sis failed
21 The module detected
fluid on the Sensors
before it should have.
Cartridge was probably
either pressed in the cen-
tre bursting the calibra-
tion pack or was stored in
conditions too warm or
too cold. Try another car-
tridge.
7DEOH7DVN:LQGRZPHVVDJHVIRUWKH%ORRG$QDO\VLV0RGXOH
0HVVDJHLQ
7DVN
:LQGRZ
,123RU
SURPSW
PHVVDJH
(UURU
&RGH
3RVVLEOH&DXVH &RUUHFWLYH$FWLRQ