Specifications
5-20 Troubleshooting Plug-in Modules
$JLOHQW0$996HULHV6HUYLFH*XLGH
Troubleshooting Plug-
in Modules
❏ If a "LQVWHDGRIDQXPHULF and a INOP message is displayed for the TEMP
module, perform the following actions:
3OHDVHQRWHWKDWIRUSURFHHGLQJWRWKHQH[WVWHSVDWUDQVGXFHUPXVWEHFRQQHFWHGWRWKH
PRGXOH
6WHS Perform a Self-Test (See “Performing the Temperature Module Self-Test” on
page 5-45). If the TEMP test numeric reads exactly the listed value, proceed to the
next step, if not, replace the module.
6WHS Refer to the following table. Correct the possible cause identified by the INOP
message and verify that the INOP disappears. If the problem still exists, replace the
module.
7DEOH,1230HVVDJHVIRUWKH7HPS0RGXOH
,1230HVVDJH 3RVVLEOH&DXVH &RUUHFWLYH$FWLRQ
<Tx> UNPLUGGED TEMP module is unplugged from the
rack.
Perform Step 5 on page 5-19.
<Tx> NO TRANSDUC Temperature probe disconnected from
the module.
Connect Temperature probe to
module.
<Tx> OVERRANGE Temperature out of range <-1
o
C or
>45
o
C.
Replace Temperature probe.
If problem still persists, replace
module.
<Tx> EQUIP MALF Malfunction in the TEMP module hard-
ware
Replace module.