module failures. The following table describes the messages and corrective actions
to take.
❏ If a VPDOO"LQDGGLWLRQWRDQXPHULF and one or more INOP messages are
displayed for the C.O. module, the values are questionable. Correct the possible
cause identified by the INOP message and verify that the INOP disappears.
7DEOH7DVN:LQGRZPHVVDJHVIRUWKH&20RGXOH
7DVN:LQGRZ
0HVVDJH
3RVVLEOH&DXVH &RUUHFWLYH$FWLRQ
no Tinj transducer
connected
Cable of Injectate
Temperature
Probe is not con-
nected or broken.
Faulty module.
Make sure that cable of the Injectate Temperature Probe is
properly connected to the C.O. Interface Cable and the con-
nector is free from dirt.
Connect Service Tool M1012-61601 to C.O. Interface
Cable
1
.
You must get the following results, otherwise the module or
C.O. Interface Cable are defect and have to be replaced:
Tinj numeric reads exactly 0.0
o
C
No prompt message “Tinj transducer connected” appears in
the Measure C.O. Task Window.
If you get the correct results from the Service Tool and the
Task Window message still appears, replace the Injectate
Temperature Probe.
1. For correct use of the Service Tool, see “Performance Test Cardiac Output” on
page 2-9
Unknown catheter Faulty C.O. Inter-
face Cable, mod-
ule or catheter.
Make sure that catheter and C.O. Interface Cable are prop-
erly connected and the connector is free from dirt.
Connect Service Tool M1012-61601 to C.O. module
a
.
If Task Window message now has disappeared, replace cath-
eter.
If Task Window message still appears, replace C.O. Inter-
face Cable.
If Task Window message still appears, replace module.
7DEOH,1230HVVDJHVZLWKTXHVWLRQDEOHYDOXHVIRUWKH&20RGXOH
,1230HVVDJH 3RVVLEOH&DXVH &RUUHFWLYH$FWLRQ
SVR MISSING
CVP
No valid CVP
available at the
moment. CVP is
automatically set to
0 mmHg for the
continuous SVR
calculation
If you need an accurate continuous SVR measurement,
perform a CVP measurement and make sure that the pres-
sure which is used for the CVP measurement is labeled
“CVP”.
Make sure you get valid numerics for CVP.
For more information refer to the Pressure Module section
in the 8VHU¶V5HIHUHQFH0DQXDO.