Specifications

5-4 Troubleshooting Plug-in Modules
$JLOHQW0$996HULHV6HUYLFH*XLGH
Troubleshooting Plug-
in Modules
 Replace the rack and plug in the tcpO
2
/tcpCO
2
module into the new rack. If the
LED on the module does not light up, replace the rack interface card in the
function box.
If a "LQVWHDGRIDQXPHULF and one or more INOP messages are displayed for the
tcpO
2
/tcpCO
2
module, perform the following actions:
3OHDVHQRWHWKDWIRUSURFHHGLQJWRWKHQH[WVWHSVDWUDQVGXFHUPXVWEHFRQQHFWHGWRWKH
PRGXOH
6WHS Perform a module self-test (See Performing the tcpO2/tcpCO2 Module Self-Test
on page 5-44). If the tcpO
2
test numeric reads exactly 60 mmHg (8.0 kPa) and the
tcpCO2 test numeric reads exactly 40 mmHg (5.3 kPa), proceed to the next step, if
not, replace the module.
6WHS Enter the tcpO
2
/tcpCO
2
Task Window by pressing the setup key on the tcpO
2
/
tcpCO
2
module. The Task Window displays messages dealing with the tcpO
2
/
tcpCO
2
module failures. The following table describes the messages and corrective
actions to take.
7DEOH7DVN:LQGRZPHVVDJHVIRUWKHWFS2
WFS&2 0RGXOH
Message Possible Cause Corrective Action
TC Equip Malf, with
prompt: module mal-
function.
Module is faulty.
Transducer has been
inserted and removed
into/from calibration
chamber too frequently.
Microswitch in Calibra-
tion chamber is intermit-
tent.
Unplug then re-plug the Module. if the
error prompt reappears within 10 sec-
onds, exchange the Module.
Unplug, then re-plug the module.
Unplug and re-plug the module. If the
error prompt reappears any time after 10
seconds, replace the Calibration cham-
ber.
TC Equip Malf, with
prompt: equipment
malfunction. The
transducer is
plugged in.
Polarization voltage or
transducer supply volt-
age is out of range.
Heating malfunction in
the transducer.
Heating control in mod-
ule is defective.
Disconnect the transducer. If the prompt
changes to module malfunction,
exchange the module. Otherwise replace
the transducer.
Swap transducers to verify prompt mes-
sage. If necessary, replace transducer.
Swap transducers to verify prompt mes-
sage. If necessary, exchange module.
No numeric values,
and no -?- in stan-
dard display when
transducer is
plugged in, but not
inserted in the Cali-
bration chamber.
Microswitch in Calibra-
tion chamber is faulty.
Replace the Calibration chamber.