Specifications
5-4 Troubleshooting Plug-in Modules
$JLOHQW0$996HULHV6HUYLFH*XLGH
Troubleshooting Plug-
in Modules
Replace the rack and plug in the tcpO
2
/tcpCO
2
module into the new rack. If the
LED on the module does not light up, replace the rack interface card in the
function box.
❏ If a "LQVWHDGRIDQXPHULF and one or more INOP messages are displayed for the
tcpO
2
/tcpCO
2
module, perform the following actions:
3OHDVHQRWHWKDWIRUSURFHHGLQJWRWKHQH[WVWHSVDWUDQVGXFHUPXVWEHFRQQHFWHGWRWKH
PRGXOH
6WHS Perform a module self-test (See “Performing the tcpO2/tcpCO2 Module Self-Test”
on page 5-44). If the tcpO
2
test numeric reads exactly 60 mmHg (8.0 kPa) and the
tcpCO2 test numeric reads exactly 40 mmHg (5.3 kPa), proceed to the next step, if
not, replace the module.
6WHS Enter the tcpO
2
/tcpCO
2
Task Window by pressing the setup key on the tcpO
2
/
tcpCO
2
module. The Task Window displays messages dealing with the tcpO
2
/
tcpCO
2
module failures. The following table describes the messages and corrective
actions to take.
7DEOH7DVN:LQGRZPHVVDJHVIRUWKHWFS2
WFS&2 0RGXOH
Message Possible Cause Corrective Action
TC Equip Malf, with
prompt: module mal-
function.
Module is faulty.
Transducer has been
inserted and removed
into/from calibration
chamber too frequently.
Microswitch in Calibra-
tion chamber is intermit-
tent.
Unplug then re-plug the Module. if the
error prompt reappears within 10 sec-
onds, exchange the Module.
Unplug, then re-plug the module.
Unplug and re-plug the module. If the
error prompt reappears any time after 10
seconds, replace the Calibration cham-
ber.
TC Equip Malf, with
prompt: equipment
malfunction. The
transducer is
plugged in.
Polarization voltage or
transducer supply volt-
age is out of range.
Heating malfunction in
the transducer.
Heating control in mod-
ule is defective.
Disconnect the transducer. If the prompt
changes to module malfunction,
exchange the module. Otherwise replace
the transducer.
Swap transducers to verify prompt mes-
sage. If necessary, replace transducer.
Swap transducers to verify prompt mes-
sage. If necessary, exchange module.
No numeric values,
and no -?- in stan-
dard display when
transducer is
plugged in, but not
inserted in the Cali-
bration chamber.
Microswitch in Calibra-
tion chamber is faulty.
Replace the Calibration chamber.