Using
Support Functions
Additional lines are displayed below the device line when a device is assigned to a central
station (as in the previous example). This does not signify that the central station is physically
connected to the same branch as the device.
When a device is assigned to a central station, the letter C or R is displayed between the
UNIT and REL columns for the central station.
A C indicates that the device is also assigned to an arrhythmia system.
An R indicates that recordings can be made at the central station.
&ROXPQ5(/
Displays the protocol release of each device on each branch.
&ROXPQ(55
Lists hexadecimal error codes for each device on the network. If the code is 00, no errors
are present. Otherwise, interpret the error codes as follows:
Convert the two-digit hexadecimal error code to an eight-digit binary code.
Compare the resulting binary code with the error code chart below: Note the extreme left
digit of the binary code is digit number 7, the second from the left is number 6, and
so on. Each 1 in the binary code corresponds to an error.
7DEOH%LQDU\&RGH(UURU&KDUW
'LJLWV (UURU,QGLFDWHG
7 6 5 4 3 2 1 0 ;
1 . . . . . . . False SPS Event Flag
1 . . . . . . Fatal Control Error Flag
1 . . . . . Recoverable Control Flag
Error
1 . . . . Physical Data Error Flag
1 . . . Logical Data Error Flag
1 . . Poll Cycle Overflow Flag
1 . Command Response Fail Flag
1 System Cycle Failure Flag