Specifications
$JLOHQW0$996HULHV6HUYLFH*XLGH
8VLQJ6XSSRUW )XQFWLRQV
Using
Support Functions
+DUGZDUH5HYLVLRQ
• Subsystem (for example: System)
• Part Number (hardware - for example: M1204-60107)
• Firmware Revision (for example: 0)
6RIWZDUH5HYLVLRQ
The software revision screen lists the software revision of the monitor and the part number of
the EEPROM.
6XPPDU\RIWKH3RZHU2Q0RGHV
The type of Power On selected by the instrument depends on the down-time of the unit as
follows:
+RW6WDUW
The instrument makes a hot start when it has been powered down for a period less than one
minute.
The active user default values, including parameter module settings are recovered and
retained on restart.
:DUP6WDUW
During a warm start the System behaves as during a cold start, but an abbreviated version of
the self-test is carried out.
The system database for patient data management is checked and recovered. All active
settings are lost.
7DEOH3RZHU2Q0RGHV$FFRUGLQJWR'RZQ7LPH
6\VWHP'RZQ7LPH 3RZHU2Q0RGH
<1 minute Hot Start
< 3 hours Warm Start
> 3 hours Cold Start
NA Extended Test can only be selected in the Service Mode.
NA X-Start (occurs when changing Operating Modes and
configuration sets).