Remove the Front Housing (see “Removing the Front Housing Assembly” on page 9-2).
If the LEFTMOST LED (AC Power) was extinguished, check fuses F1and F2 (AC
fuses), and F6 (18V fuse). Fuses F1 and F2 can be checked visually. Fuse F6 must be
checked with an ohmmeter and is considered good if it reads a short. If the fuses are
good, verify that the monitor was plugged into a sufficient power source. If it was,
replace the power supply assembly (see “Chapter 8 Repairing the Agilent Model V24CT/
V26CT”).
If either F1 or F2 are blown, replace them (see Chapter 11 for part number).