User`s guide
Chapter 3 79
Setting Measurement Conditions
Setting Stimulus Conditions
3. Setting Measurement
Conditions
Figure 3-7 Sweep Time and Sweep Delay Time
Sweep delay is time before starting a sweep for each stimulus (source) port
Setting Up the Sweep Time (Manual Sweep Time Mode)
Step 1. Press or to select the channel for which sweep time will be set.
Step 2. Press .
Step 3. Press
Sweep Time.
Step 4. Using the ENTRY block keys on the front panel, input the desired sweep time (in seconds).
If the previous operation mode was automatic sweep time mode, entering a new sweep
time forces the machine to switch to manual sweep time mode.
Switching to Automatic Sweep Time Mode
Step 1. Press or to select the channel which will be switched to
automatic sweep time mode.
Step 2. Press .
Step 3. Press
Sweep Time.
Step 4. Press .
By entering zero (seconds), automatic sweep time becomes effective.