756 Appendix F
Replacing the 8753ES with the E5070B/E5071B
Comparing Functions
Reading/Writ
ing data
Transfer format Intra-device binary
IEEE 32-bit floating point
IEEE 64-bit floating point
ASCII
PC-DOS 32-bit floating point
IEEE 32-bit floating point
IEEE 64-bit floating point
ASCII
PC-DOS 32-bit floating point
Object to be read/written Raw data array
Calibration coefficient data array
(before interpolating)
Corrected data array
Formatted memory array
Memory trace
Calibration kit array data
Power meter calibration coefficient
array (before interpolating)
Calibration coefficient data array
(after interpolating)
Corrected data array
Corrected memory array
Formatted data array
Formatted memory array
Power meter calibration coefficient
array (after interpolating)
Object to be read only Pre-raw data (in Take4 mode)
Calibration coefficient array (after
interpolating)
Power meter calibration coefficient
array (after interpolating)
Reading data for a designated point or scope Allowed only for data traces Not allowed
Marker Number of markers Up to five markers per channel Nine markers per tracer (exclude
reference marker)
Active marker One marker per channel One marker per trace
Delta marker (Reference marker) Designates the marker displayed as
the reference marker.
You can designate the reference
marker independently from the
markers displayed.
Marker coupling Setting of Couple/Uncouple
definable
Setting of Couple/Uncouple
definable
(coupling available only between
markers on the same channel)
Marker move mode Continuous/discrete
(definable channel by channel)
Continuous/discrete
(definable trace by trace)
Fixed marker One marker definable per channel Not available
Marker value display On/Off setting for the marker value
display definable per channel
All marker values for active traces
are displayed on each channel.
*2
Display/No Display setting for the
marker table displaying all marker
values definable
Table F-5 Functions of the 8753ES vs. the E5070B/E5071B
Function 8753ES E5070B/E5071B