User`s guide

Appendix F 755
Replacing the 8753ES with the E5070B/E5071B
Comparing Functions
F. Replacing the 8753ES
with the E5070B/E5071B
Calibration
(cont'd.)
Defining the
calibration kit
Calibration class S11A (OPEN)
S11B (SHORT)
S11C (LOAD)
S22A (OPEN)
S22B (SHORT)
S22C (LOAD)
Forward Transmission
Forward Match
Reverse Transmission
Reverse Match
Response
Response & Isolation
TRL Thru
TRL Reflect
TRL Line/Match
OPEN
SHORT
LOAD
THRU
(OPEN, SHORT, and LOAD are set
for each port; THRU is set between
ports)
TRL Thru
TRL Reflect
TRL Line/Match
Class indication label Editable Not editable
Number of standards that can be
registrated in calibration classes
Up to seven types One type
Type of standard OPEN, SHORT, LOAD, and THRU
Arbitrary Impedance
OPENOPEN, SHORT, LOAD, and
THRU
Arbitrary Impedance, None
Standard
parameters
Common to
standards
Offset delay
Offset loss
Offset characteristic impedance
Frequency range
Offset type (coaxial, waveguide)
Unique to
OPEN
C0, C1, C2, C3
Unique to
SHORT
Not available L0, L1, L2, L3
Unique to
LOAD
Type of standard (fixed, sliding,
offset)
Not available (the type of standard is
treated as fixed load.)
Unique to
THRU
Not available
Arbitrary
Impedance
Type of standard (fixed, sliding)
Arbitrary Impedance
Arbitrary Impedance
Standard label Editable
ECal Type of calibration Enhanced response calibration
S11 1-port calibration
S22 1-port calibration
Full 2-port calibration
Enhanced response calibration
Response (thru) calibration
Full 1-port calibration
Full 2-port calibration
Full 3-port calibration (Opts.
313/314/413/414)
Full 4-port calibration (Opts.
413/414)
Omission of isolation Allowed
Manual THRU measurement Allowed Not allowed
Reading the frequency array Allowed Not allowed
Reading the module information Allowed Not allowed
Table F-5 Functions of the 8753ES vs. the E5070B/E5071B
Function 8753ES E5070B/E5071B