User`s guide
668 Appendix D
Softkey Functions
Calibration Menu
(Continued)
Calibrate
Response (Thru)
Displays softkeys for selecting options for response calibrations using the THRU standard.
Select Ports Displays softkeys for selecting test ports.
2-1 (S21) Selects the transmission test (measurement of S
21
) for test port
1→2.
:SENS{1-16}:CORR:COLL:METH:THRU 2,1
3-1 (S31)
*1
Selects the transmission test (measurement of S
31
) for test port
1→3.
:SENS{1-16}:CORR:COLL:METH:THRU 3,1
4-1 (S41)
*2
Selects the transmission test (measurement of S
41
) for test port
1→4.
:SENS{1-16}:CORR:COLL:METH:THRU 4,1
1-2 (S12)
Selects the transmission test (measurement of S
12
) for test port
2→1.
:SENS{1-16}:CORR:COLL:METH:THRU 1,2
3-2 (S32)
*1
Selects the transmission test (measurement of S
32
) for test port
2→3.
:SENS{1-16}:CORR:COLL:METH:THRU 3,2
4-2 (S42)
*2
Selects the transmission test (measurement of S
42
) for test port
2→4.
:SENS{1-16}:CORR:COLL:METH:THRU 4,2
1-3 (S13)
*1
Selects the transmission test (measurement of S
13
) for test port
3→1.
:SENS{1-16}:CORR:COLL:METH:THRU 1,3
2-3 (S23)
*1
Selects the transmission test (measurement of S
23
) for test port
3→2.
:SENS{1-16}:CORR:COLL:METH:THRU 2,3
4-3 (S43)
*2
Selects the transmission test (measurement of S
43
) for test port
3→4.
:SENS{1-16}:CORR:COLL:METH:THRU 4,3
1-4 (S14)
*2
Selects the transmission test (measurement of S
14
) for test port
4→1.
:SENS{1-16}:CORR:COLL:METH:THRU 1,4
2-4 (S24)
*2
Selects the transmission test (measurement of S
24
) for test port
4→2.
:SENS{1-16}:CORR:COLL:METH:THRU 2,4
3-4 (S34)
*2
Selects the transmission test (measurement of S
34
) for test port
4→3.
:SENS{1-16}:CORR:COLL:METH:THRU 3,4
Cancel
Returns to the softkey display screen one level higher.
Thru Executes a THRU response calibration on the test ports selected in
the
Select Ports menu. This calibration is effective for eliminating
transmission tracking errors.
:SENS{1-16}:CORR:COLL:THRU
Isolation
(Optional)
Executes an isolation calibration on the test ports selected in the
Select Ports menu. This calibration is effective for eliminating
isolation errors.
:SENS{1-16}:CORR:COLL:ISOL
Done
Terminates a calibration process and calculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:SAVE
Cancel
Displays softkeys for canceling a calibration.
OK Cancels a calibration process and clears the measurement data of
standard.
:SENS{1-16}:CORR:COLL:CLE
Cancel
Returns to the softkey display screen one level higher.
Return Returns to the softkey display screen one level higher.
Enhanced
Response
Displays softkeys for executing enhanced response calibrations.
Select Port Displays softkeys for selecting a test port.
2-1 (S21 S11) Selects the transmission test (measurement of S
21
) for test port
1→2.
:SENS{1-16}:CORR:COLL:METH:ERES 2,1
Key Operation Function SCPI Command