User`s guide

Chapter 18 583
Information on Maintenance
Removing/Mounting Removable Hard Disk
18. Information on
Maintenance
Writing System Correction Data from Backup Memory to Removable
Hard Disk
The following procedure shows how to write the system correction data stored in the
backup memory to the newly mounted removable hard disk.
Step 1. Turn on the E5070B/E5071B.
Step 2. Press .
Step 3. Press
Load Project... in the softkey menu.
Step 4. A dialog box appears for you to select the program to be loaded. Select
RestoreSysCorFile.vba from the D:\Agilent\Service folder and then press the
Open button.
Step 5. Press .
Step 6. The RestoreSysCorFile dialog box appears (Figure 18-3). Click
OK.
Figure 18-3 RestoreSysCorFile Dialog Box
CAUTION Once the program has terminated, restart the E5070B/E5071B to enable the system
correction data.