User`s guide

Chapter 16 539
Specifications and Supplemental Information
Corrected System Performance
16. Specifications and
Supplemental Information
Corrected System Performance
The specifications in this section apply for measurements made with the Agilent
E5070B/E5071B Network Analyzer under the following conditions:
No averaging applied to data
Environmental temperature of 23°C ±5°C, with less than 1°C deviation from the
calibration temperature
Response and isolation calibration not omitted
Table 16-1 System Dynamic Range
Description Specification Supplemental Information
System Dynamic Range
*1*2
300 kHz to 3 MHz IF bandwidth = 3 kHz 85 dB
3 MHz to 1.5 GHz 95 dB 98 dB
1.5 GHz to 3 GHz 97 dB 100 dB
3 GHz to 4 GHz 96 dB 99 dB
4 GHz to 6 GHz 92 dB 94 dB
6 GHz to 7.5 GHz 87 dB 90 dB
7.5 GHz to 8.5 GHz 80 dB 83 dB
300 kHz to 3 MHz IF bandwidth = 10 Hz 110 dB
3 MHz to 1.5 GHz 120 dB 123 dB
1.5 GHz to 3GHz 122 dB 125 dB
3 GHz to 4 GHz 121 dB 124 dB
4 GHz to 6 GHz 117 dB 119 dB
6 GHz to 7.5 GHz 112 dB 115 dB
7.5 GHz to 8.5 GHz 105 dB 108 dB
*1.The test port dynamic range is calculated as the difference between the test port rms noise floor and the source
maximum output power. The effective dynamic range must take measurement uncertainty and interfering signals
into account.
*2.May be limited to 90 dB at particular frequencies below 350MHz or above 4.25GHz due to spurious receiver
residuals.