User`s guide

534 Chapter 15
Measurement Examples
Executing Power Calibration
2. Setting Address of Power Meter
Follow these steps to configure the power meter's GPIB address.
3. Setting Stimulus Condition
Follow these steps to configure the stimulus conditions:
4. Executing Zero Adjustment and Calibration of Power Sensor
Execute the zero adjustment and calibration of the power sensor according to the E4418B
Power Meter User's Guide.
5. Setting Calibration Data Measurement Conditions
Follow these steps to set the calibration data measurement conditions:
Setting Description Key Operation
GPIB address of the power meter: 14
-
Misc Setup - GPIB Setup - Power Meter
Address
-
Setting Description Key Operation
Presetting
-
OK
Sweep type: Power
-
Sweep Type - Power Sweep
Fixed frequency: 1 GHz
-
Power - CW Freq -
Power range: -40 dBm to -8 dBm
- Power - Power Ranges - -40 to -8
Start value: -40 dBm
Stop value: -10 dBm
Number of points: 61
-
Points -
Setting Description Key Operation
Selecting the test port: 1
-
Power Calibration - Select Port - 1
Selection of power sensor: A
-
Power Calibration - Use Sensor [ A ]
Number of measurements at one
measurement point: 4
-
Power Calibration - Num of Readings -
Tolerance during power calibration: 5 dB
- Power Calibration - Tolerance -