User`s guide

500 Chapter 15
Measurement Examples
Evaluating a Duplexer
Figure 15-9 EcalAssistant (port/channel selection) dialog box
Step 5. Following the instructions in the dialog box, select the type of ECal, test ports, and the
channel.
Step 6. Press the
Next button. The EcalAssistant (connection) dialog box (Figure 15-10) appears.
Figure 15-10 EcalAssistant (connection) dialog box
Step 7. Connect the ECal module to the test ports according to the instructions in the dialog box
and then press
Measure.
Setup Description Key Operation
Select Ports
ECal type: Full 3-port calibration
3 Port
Test ports to be used for ECal: 1, 2, and 3 1, 2, 3
Select Channel
Channel on which ECal is to be executed:
Channel 1
Channel: 1
Setup Description Key Operation
Executing calibration between ports 1 and 2 (After making the connections shown in Figure
15-11 (a)), press
Measure - Next
Executing calibration between ports 1 and 3 (After making the connections shown in Figure
15-11 (b)), press Measure - Next
Executing calibration between ports 2 and 3 (After making the connections shown in Figure
15-11 (c)), press Measure - Next